如何達成單一軟式輸尿管鏡完成執行200例以上之經驗分享
 
陳鴻毅、吳俊德、黃賢祥、林政鋒、陳文祥、林承家
基隆長庚醫院 外科部 泌尿科
How to achieve 200+ cases durability with flexbile ureteoscopes: a singel hospital experience
 
Hung-Yi Chen, Chun-Te Wu, Shian-Shiang Huang, Cheng-Feng Lin, Wen-Hsiang Chen, Cheng-chia Lin
 
 Divisions of Urology, Department of Surgery, Chang Gung Memorial Hospital, Keelung
 
Purpose:
  To describe a novel operative method for extended durability and improved ease of use of current flexible ureteroscopes.
 
Material and methods:
328 patients underwent retrograde intrarenal surgery at our hospital. The modified method was used by a single surgeon. The outcome and durability for flexible ureteroscopes were assessed.
 
Results:
An Olympus digital flexible URF-V ureteroscope was used, and it required repair thrice during the study period. During factory maintenance, distal working channel damage was noted twice and outer bending rubber damage was noted once. Despite the damage and repairs, the deflection system was almost entirely intact after high-frequency use.
 
Conclusion:
  The durability of a flexible ureteroscope (FURS) determines the efficacy of retrograde intrarenal surgery for renal stones. In this report, we described our modified upside-down technique for manipulation of the FURS in unequal dual deflection in order to preserve the deflection apparatus, which yielded greatly prolonged durability. Additionally, the use of mimic drive turning has decreased the time needed to train urologists.